ICT / ITE / Medical Power Supply

GTM964500PWWWVV.V-F Image
Model ImageGTM964500PWWWVV.V-F, ICT / ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals: GTM964500PWWWVV.V-F, ICT / ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals:
Information
Model NumberGTM964500PWWWVV.V-F
DescriptionGTM964500PWWWVV.V-F, ICT / ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals: GTM964500PWWWVV.V-F, ICT / ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals:
Agency Documents
CE EC-Declaration for Medicalhttps://globtek.secure.force.com/models/ec_declaration?id=a0O3a00000Ly9m5EAB
CE EC-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ec_declaration?id=a0O3a00000Ly9m5EAB
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for Medicalhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O3a00000Ly9m5EAB
UKCA-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O3a00000Ly9m5EAB
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://globtek.secure.force.com/models/Morocco_SDoC_Declaration?id=a0O3a00000Ly9m5EAB
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0O3a00000Ly9m5EAB
Prop 65 Statementhttps://en.globtek.com/prop-65-statement.pdf
REACH Declarationhttps://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttps://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypeOpen Frame/Internal / Per installazione interna, versione 'Open Frame'
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor / Alimentatore AC-DC di tipo switching
CategoryICT / ITE / Medical Power Supply / ICT / ITE / Medical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)6A
Max Output Power Rating (W)450
Output Voltage Range (V)12-54
Auxiliary/Fan Output Voltage (V)12
Auxiliary/Fan Output Power (W)15W Forced Air, 10W Convection
EfficiencyUSA DOE Level VI / Eco-design Directive 2009/125/EC, (EU) 2019/1782
Ingress ProtectionN/A
Size (mm)127.0 (L) x 76.2 (W) x 38.5 (H)
EnclosureUser-added image

Customer may need to provide EMI shielding if not in metal chassis
User-added image
Rating Table
Model Number (Final ordering #/Part No to be defined)VoltageAmps (Convection)Amps (Forced Air)Watts (Convection)Watts (Forced Air)RFQ
* (Final ordering #/Part No to be defined)
Specifications

Features

  • High power density 3" x 5"� format
  • Auxiliary +12V fan output
  • Auxiliary +5V standby output
  • Power-good, remote-off, & remote-sense
  • Class I, Class II, or Class II with Functional Earth input configurations
  • ITE/ICT and medical (2 x MOPP) certified

Input

Input Voltage:Specified: 90-264VAC, Nameplate: 100-240VAC
Input Frequency:Specified: 47-63Hz, Nameplate: 50-60Hz
Input Current:5.8A typ. @ 90VAC, 2.3A typ. @ 230VAC (full load)
No Load Input Power:<150mW
Inrush Current:<35A @ @ 100VAC , <70A @ 230VAC (cold start @ 25°C)
Efficiency:85 - 95% typ. (full load)
Power Factor:>0.95 @ 115VAC, >0.90 @ 230VAC (full load)

Output

Output Power:For 12-18.9V output: 400W (25CFM forced air), 225W (convection), inclusive of fan/aux power
For 19-54.0V output: 450W (25CFM forced air), 250W (convection), inclusive of fan/aux power
Turn-on Delay:< 3 seconds @ 90VAC, < 2 seconds @ 115VAC
Output Voltage Tolerance:± 2.0% (worst case, set point + load/line regulation)
Load Regulation:± 0.5% (measured at connector)
Line Regulation:± 0.2% (measured at connector)
Ripple:1% pk-pk (using a 47µF low-ESR cap + 0.1µF ceramic capacitor, measured @ 20MHz BW)
Transient Response:<5% deviation (with 25% load step), recovery to within 1% in 1ms
Startup Overshoot:<5% (full load)
Hold-up Time:8ms typ. (full load)
+12V Fan Output:800mA (max)
+5V Standby Output:1A (max)

Protections

Input Protection:MOV transient suppressor, dual fuses (line + neutral)
Over-Voltage Protection:110-150%. Latched-off, cycle AC to reset
Over-Current Protection:105-150% typ. Non-latched with auto-recovery
Short Circuit Protection:Non-latched with auto-recovery
Over-Temperature Protection:Latched-off, cycle AC to reset

Environmental

MTBF:200,000 hours @ 25°C ambient
Operating Temperature:-20°C to 40°C (Full output power), -40°C to 80°C (Derated)
Storage Temperature:-40°C to 85°C
Humidity:0% to 95% relative humidity, non-condensing
Altitude:5000m max.
Vibration:Conforms to EN60068-2-6

Safety

Dielectric Withstand Voltage:4000VAC or 5656VDC from input to output
3000VAC or 4242VDC from input to earth
Intended Medical Application:Type-B or Type-BF
Touch Current:NC: <100µA SFC: <500µA
Earth Leakage CurrentNC: <0.5mA SFC: <1.0mA
Means of Protection:2 x MOPP
Primary-Secondary Bridging Capacitance:Two IEC60384-14 Y1 capacitors in series

EMC

Applicable Standards:Medical: EN60601-1-2, 4th edition
Emissions: EN55032, EN 55011, EN61000-6-3, EN61000-6-4, CISPR11/CISPR22
Immunity: EN55024, EN61000-6-1, EN61000-6-2
Conducted Emissions:Class B, FCC Part 15, Class B (with resistive load)
Radiated Emissions:Class B, FCC Part 15, Class B (with resistive load)
Harmonic Current Voltage Distortion:EN61000-3-2, Class A
Voltage Fluctuations/Flicker:EN61000-3-3
Electrostatic Discharge (ESD) Immunity:EN61000-4-2, 10KV contact Discharge, 20KV air discharge (applied to input or output terminals)
Radiated RF Immunity:EN61000-4-3, 10V/m 80-2700MHz, 80% 1KHz AM.
EFT/Burst Immunity:EN61000-4-4, 4KV/100kHz.
Line Surge Immunity:EN61000-4-5, 2KV differential, 4KV common-mode
Conducted RF Immunity:EN61000-4-6, 3VRMS, 80% 1KHz AM
Power Frequency Magnetic Field Immunity:EN61000-4-8, 30A/m
Voltage Dip Immunity:EN61000-4-11

 

Derating CurveUser-added image
User-added image
Qty (ea.)/Carton
Individual Box/Gift Box Size (mm)
Individual Box/Gift Box Color
Master Carton Size (mm)TBD
Gross Weight (kg)
Label Width (mm)TBD
Label Height (mm)TBD
Input Configuration
DescriptionUser-added image
Optional Input Wiring Kits are available:
PN/Ordering NoLengthEnd AEnd B
C810070M6VAR2(R)60IEC 60320 C8JST VAR-2
C1410070M6VAR2(R)60IEC 60320 C13JST VAR-2
VAR2210070M7VAR2(R)73JST VAR-2JST VAR-2
Custom lengths, construction, cable types, ferrites, filters, and termination available upon request.
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
Conforms to UL STD. 62368-1
Certified to CSA STD C22.2 NO.62368-1
Conforms to UL STD. 62368-1 Certified to CSA STD C22.2 NO.62368-1
User-added image
 
Morocco SDoC declaration http://www.globtek.info/certs/Morocco%20SDoC%20Declaration/
User-added image
 
UKCA Certification
Popular Products (Stock items)
SKUVolts OutCurrent Out (Amps)ConnectorRFQSample Request

Cos’è un alimentatore AC-DC di tipo open frame ?

Anche noti come convertitori AC-DC, si tratta di una tipologia di alimentatori per uso da interno. Gli alimentatori AC-DC open frame sono spesso integrati in altri sistemi o alloggiati in particolari case metallici ed utilizzati per convertire la tensione da AC a DC.

Cos’è un alimentatore conforme alla norma IEC 62368-1:2018 ?

Tale norma fissa i requisiti di sicurezza per sistemi elettrici ed elettronici in ambito audio, video, telecomunicazioni e applicazioni per uffici e similari per tensioni non superiori a 600 V. I requisiti fissati riguardano i dispositivi di protezione e la loro effettiva implementazione nel sistema allo scopo di prevenire danni a persone o cose per elettrocuzione o incendio. Un alimentatore conforme alla IEC 62368-1 è pertanto un convertitore AC/DC che ha superato i test previsti da tale norma.

Gli alimentatori IEC 62368-1 possono essere considerati come succedanei di quelli conformi IEC 60950-1?

È già in corso la transizione delle norme IEC 60065 e 60950-1 verso la nuova IEC 62368-1. In Nord America le certificazioni ETL e UL 60950-1 continueranno a rimanere valide, a meno di ulteriori novità.

Vi sono differenze sostanziali fra la IEC 60950-1 e la IEC 62368-1 per gli alimentatori?

Sebbene l’impostazione generale della nuova norma appaia diversa, di fatto la maggior parte dei prodotti che risultavano conformi alla IEC 60950-1 lo sono anche per la IEC 62368-1 apportando minime modifiche o, in taluni casi, senza alcuna modifica necessaria. In ogni caso, GlobTek continuerà a produrre alimentatori IEC 60950-1 a meno di ulteriori novità.

Cos’è un alimentatore conforme alla norma IEC 62368-1:2018 ?

Tale norma fissa i requisiti di sicurezza per sistemi elettrici ed elettronici in ambito audio, video, telecomunicazioni e applicazioni per uffici e similari per tensioni non superiori a 600 V. I requisiti fissati riguardano i dispositivi di protezione e la loro effettiva implementazione nel sistema allo scopo di prevenire danni a persone o cose per elettrocuzione o incendio. Un alimentatore conforme alla IEC 62368-1 è pertanto un convertitore AC/DC che ha superato i test previsti da tale norma.

Gli alimentatori IEC 62368-1 possono essere considerati come succedanei di quelli conformi IEC 60950-1?

È già in corso la transizione delle norme IEC 60065 e 60950-1 verso la nuova IEC 62368-1. In Nord America le certificazioni ETL e UL 60950-1 continueranno a rimanere valide, a meno di ulteriori novità.

Vi sono differenze sostanziali fra la IEC 60950-1 e la IEC 62368-1 per gli alimentatori?

Sebbene l’impostazione generale della nuova norma appaia diversa, di fatto la maggior parte dei prodotti che risultavano conformi alla IEC 60950-1 lo sono anche per la IEC 62368-1 apportando minime modifiche o, in taluni casi, senza alcuna modifica necessaria. In ogni caso, GlobTek continuerà a produrre alimentatori IEC 60950-1 a meno di ulteriori novità.

Cos’è il livello VI di efficienza?

L’ordinamento numerico da I a VI si riferisce alla marcatura internazionale di prodotti in relazione al livello di efficienza di un certo prodotto. I requisiti previsti per la classificazione si basano sulla misura dell’efficienza durante il funzionamento e sulla valutazione del consumo a vuoto e a diversi livelli di carico. Il consumo a vuoto viene registrato in assenza di carico o con carico disconnesso. Il livello di efficienza VI è ormai obbligatorio in diverse nazioni per specifici prodotti o per specifiche applicazioni. La Direttiva Europea Eco-design 2009/125/EC ha di fatto inserito, per i Paesi Europei, i requisiti del livello di efficienza VI nell’ambito delle normative che regolano la vendita e l’importazione di alimentatori per uso esterno.

Come funziona un adattatore AC-DC?

L’adattatore AC-DC è un convertitore da esterno integrato all’interno di un piccolo case costruito attorno ad una spina AC per collegamento alla rete, da cui derivano anche i vari termini plug, wall plug in, plug-in o plugtop. Una volta collegato alla rete elettrica AC, la tensione viene convertita in DC per l’alimentazione del carico attraverso il cavo di uscita.

Richiesta informazioni
Call
Email
💬
Text
💁
Chat
?
Richiedi un preventivo
Ti possiamo aiutare?
Ti possiamo aiutare?