Preferred ratings shown below. Custom output voltages available in 0.1V increment from the minimum voltage listed to the maximum voltage listed; contact sales.
Model Number (Final ordering #/Part No to be defined)
Voltage
Amps (A)
Watts (W)
RFQ
GTM-43001130505*
4.5
2.2
9.9
GTM-43001130505-XXXX
CUSTOM
CUSTOM
CUSTOM
* (Final ordering #/Part No to be defined)
Specifications
Input
Input Voltage:
Specified: 90-264VAC, Nameplate: 100-240VAC 100% rated load current for 90-264VAC 85% rated load current for 85-264VAC 100% rated load current for 110-370VDC
Input Frequency:
Specified: 47-63Hz, Nameplate: 50-60Hz
No Load Input Power:
<75mW @ 230VAC (EU CoC Tier 2 compliant)
Inrush Current:
40A maximum, cold start @ 230VAC input
Efficiency:
DoE Level VI compliant, EU CoC Tier 2 compliant
Output
Turn-on Delay:
1 second max. @ nominal line and full load
Output Regulation
± 2% (measured at the output connector)
Line Regulation:
± 1% (measured at the output connector, full load)
-20°C to 55°C (full load) See derating table for output power capability at other temperatures. Extended low temperature range available as a custom option.
Storage Temperature:
-30°C to 80°C
Humidity:
0% to 95% relative humidity, non-condensing
Altitude:
5000m
Ingress protection:
NA
Cooling:
Convection
RoHS:
Complies with EU 2011/65/EU amended by EU 2015/863 and China SJ/T 11363-2006
Safety
Dielectric Withstand Voltage:
4000VAC or 5656VDC from input to output 3000VAC or 4242VDC from input to earth (Class I models)
Touch Current:
Class I models: < 20uA Class II models: <75uA
Earth Leakage Current:
Class I models: <300uA Class II models: N/A
Isolation Options:
Floating output or Class I earthed output (desktop models)
Earth Continuity Test:
< 0.1 ohm between AC inlet earth pin and PCB termination point
Class B, FCC Part 15, Class B (with resistive load)
Radiated Emissions:
Class B, FCC Part 15, Class B (with resistive load)
Harmonic Current Voltage Distortion:
EN61000-3-2, Class A
Voltage Fluctuations/Flicker:
EN61000-3-3
Electrostatic Discharge (ESD) Immunity:
EN61000-4-2, 8KV contact discharge, 20KV air discharge
Radiated RF Immunity:
EN61000-4-3, 10V/m 80-1000MHz, 3V/m 1-2.7GHz, 80% 1KHz AM
EFT/Burst Immunity:
EN61000-4-4, 4KV/100kHz.
Line Surge Immunity:
EN61000-4-5, 2KV differential, 4KV common-mode
Conducted RF Immunity:
EN61000-4-6, 3VRMS, 80% 1KHz AM
Power Frequency Magnetic Field Immunity:
EN61000-4-8, 30A/m
Voltage Dip Immunity:
EN61000-4-11
Enclosure
Housing:
High impact plastic, 94V0 polycarbonate, non-vented
AC Input Mechanical Options:
Wall plug-in, changeable blades (Class II only) Desktop housing, extended housing: C6, C8, C14, or C18 IEC inlet (Class I or Class II)
Markings:
Label or laser printed
Special Options
Removal of LED (optional)
Custom cord-sets (custom cord and/or connector types). Note: 5VDC USB power supplies with Micro-B or Type-C connectors have a voltage regulation specification of ±8% (4.6V to 5.4V).
Custom markings
Short term output surge capability
Reduced leakage current versions (10uA max) for Type-CF medical applications and/or for AC noise sensitive equipment. (15W or less, 5 to 24V output voltage range.)
Tightened output voltage tolerance
Reduced output voltage ripple
Improved reliability PCB laminate with plated-through-holes for IPC610 Class 2 compliance
Custom housing colors
Quasi constant-current output for SLA (lead-acid) battery charging applications
Derating Curve
Qty (ea.)/Carton
60
Individual Box/Gift Box Size (mm)
Egg Crate
Individual Box/Gift Box Color
white
Master Carton Size (mm)
480 x 450 x 205
Gross Weight (kg)
15.2
Label Width (mm)
29.8
Label Height (mm)
20.8
Input Configuration
Description
Output Configuration
Common output connector options: Click to add to your request.
Contact GlobTek for your specific requirements or custom solutions.
Popular Products (Stock items)
SKU
Volts Out
Current Out (Amps)
Connector
RFQ
Sample Request
2185153
4.5
2.2
Cos’è il livello VI di efficienza?
L’ordinamento numerico da I a VI si riferisce alla marcatura internazionale di prodotti in relazione al livello di efficienza di un certo prodotto. I requisiti previsti per la classificazione si basano sulla misura dell’efficienza durante il funzionamento e sulla valutazione del consumo a vuoto e a diversi livelli di carico. Il consumo a vuoto viene registrato in assenza di carico o con carico disconnesso. Il livello di efficienza VI è ormai obbligatorio in diverse nazioni per specifici prodotti o per specifiche applicazioni. La Direttiva Europea Eco-design 2009/125/EC ha di fatto inserito, per i Paesi Europei, i requisiti del livello di efficienza VI nell’ambito delle normative che regolano la vendita e l’importazione di alimentatori per uso esterno.
Cos’è un alimentatore AC-DC di tipo plug-in ?
Anche noti come adattatori AC o convertitori AC-DC, si tratta di una tipologia di alimentatori per uso da esterno. Alcune apparecchiature elettriche necessitano di alimentazione di potenza ma non dispongono di una circuiteria interna per poter gestire un collegamento diretto alla rete. In casi del genere, un alimentatore AC viene utilizzato per convertire la tensione AC di rete in tensione DC adatta per l’alimentazione dell’apparecchiatura in questione.
Come funziona un adattatore AC-DC?
L’adattatore AC-DC è un convertitore da esterno integrato all’interno di un piccolo case costruito attorno ad una spina AC per collegamento alla rete, da cui derivano anche i vari termini plug, wall plug in, plug-in o plugtop. Una volta collegato alla rete elettrica AC, la tensione viene convertita in DC per l’alimentazione del carico attraverso il cavo di uscita.